Blood Pressure Measurement Device Accuracy Evaluation: Statistical Considerations with an Implementation in R

Autor: Tanvi Chandel, Victor Miranda, Andrew Lowe, Tet Chuan Lee
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: Technologies, Vol 12, Iss 4, p 44 (2024)
Druh dokumentu: article
ISSN: 2227-7080
DOI: 10.3390/technologies12040044
Popis: Inaccuracies from devices for non-invasive blood pressure measurements have been well reported with clinical consequences. International standards, such as ISO 81060-2 and the seminal AAMI/ANSI SP10, define protocols and acceptance criteria for these devices. Prior to applying these standards, a sample size of N >= 85 is mandatory, that is, the number of distinct subcjects used to calculate device inaccuracies. Often, it is not possible to gather such a large sample. Many studies apply these standards with a smaller sample. The objective of the paper is to introduce a methodology that broadens the method first developed by the AAMI Sphygmomanometer Committee for accepting a blood pressure measurement device. We study changes in the acceptance region for various sample sizes using the sampling distribution for proportions and introduce a methodology for estimating the exact probability of the acceptance of a device. This enables the comparison of the accuracies of existing device development techniques even if they were studied with a smaller sample size. The study is useful in assisting BP measurement device manufacturers. To assist clinicians, we present a newly developed “bpAcc” package in R to evaluate acceptance statistics for various sample sizes.