Accuracy of handheld blood glucose meters at high altitude.

Autor: Pieter de Mol, Hans G Krabbe, Suzanna T de Vries, Marion J Fokkert, Bert D Dikkeschei, Rienk Rienks, Karin M Bilo, Henk J G Bilo
Jazyk: angličtina
Rok vydání: 2010
Předmět:
Zdroj: PLoS ONE, Vol 5, Iss 11, p e15485 (2010)
Druh dokumentu: article
ISSN: 1932-6203
DOI: 10.1371/journal.pone.0015485
Popis: BACKGROUND: Due to increasing numbers of people with diabetes taking part in extreme sports (e.g., high-altitude trekking), reliable handheld blood glucose meters (BGMs) are necessary. Accurate blood glucose measurement under extreme conditions is paramount for safe recreation at altitude. Prior studies reported bias in blood glucose measurements using different BGMs at high altitude. We hypothesized that glucose-oxidase based BGMs are more influenced by the lower atmospheric oxygen pressure at altitude than glucose dehydrogenase based BGMs. METHODOLOGY/PRINCIPAL FINDINGS: Glucose measurements at simulated altitude of nine BGMs (six glucose dehydrogenase and three glucose oxidase BGMs) were compared to glucose measurement on a similar BGM at sea level and to a laboratory glucose reference method. Venous blood samples of four different glucose levels were used. Moreover, two glucose oxidase and two glucose dehydrogenase based BGMs were evaluated at different altitudes on Mount Kilimanjaro. Accuracy criteria were set at a bias 6.5 mmol/L) and