How Much PEEP Does High Flow Deliver via Tracheostomy? A Literature Review and Benchtop Experiment

Autor: Martin Thomas, Riddhi Joshi, Grant Cave
Jazyk: angličtina
Rok vydání: 2021
Předmět:
Zdroj: Critical Care Research and Practice, Vol 2021 (2021)
Druh dokumentu: article
ISSN: 2090-1305
2090-1313
DOI: 10.1155/2021/6036891
Popis: Background. High flow tracheostomy (HFT) is a commonly used weaning and humidification strategy for tracheostomised patients, but little is known as to how much PEEP or mechanical benefit it offers. Patient anatomy and device characteristics differentiate it from high flow nasal cannula and the physiological effects observed. Objectives. (1) To review the available literature on the effects of HFT on airway pressure and indices of gas exchange. (2) To quantify PEEP generated by a HFT circuit. Methods. A randomised benchtop experiment was conducted, with a size 8 uncuffed Portex tracheostomy connected to an Optiflow™ with Airvo 2™ humidifier system. The tracheostomy tube was partially immersed in water to give rise to a column of water within the inner surface of the tube. An air fluid interface was generated with flows of 40 L/min, 50 L/min, and 60 L/min. The amount of potential PEEP (pPEEP) generated was determined by the distance the water column was pushed downward by the flow delivered. Findings. Overall 40 L/min, 50 L/min, and 60 L/min provided pPEEP of approximately 0.3 cmH2O, 0.5 cmH2O, and 0.9 cmH2O, respectively. There was a statistically significant change in pPEEP with change in flows from 40–60 L/min with an average change in pPEEP of 0.25–0.35 cmH2O per 10 L/min flow (p value
Databáze: Directory of Open Access Journals