Skin thermal injury at site of hypodermic electrodes placed for intraoperative neuromonitoring during thyroidectomy: one case report

Autor: XING Xujian, CHEN Liang, HE Wen, FEI Jian
Jazyk: čínština
Rok vydání: 2024
Předmět:
Zdroj: Waike lilun yu shijian, Vol 29, Iss 03, Pp 260-263 (2024)
Druh dokumentu: article
ISSN: 1007-9610
DOI: 10.16139/j.1007-9610.2024.03.13
Popis: In order to reduce intraoperative recurrent laryngeal nerve injury, intraoperative neural monitoring(IONM) is widely used in thyroid-related surgery, especially in reoperations. However, some incidental injuries have also been observed during IONM. This article reported a case suffered skin thermal injury at the site of hypodermic needle electrode placement during IONM in Shanghai International Medical Center. Combining relevant adverse event information and analyzing the causes was essential to understand and prevent potential risks, and prioritize patient safety.