A force sensor system was developed to give real-time visual feedback on a range of force. In a prospective observational cross-section study, twenty-two anaesthesia nurses applied cricoid pressure at a target range of 30-40 Newtons for 60 seconds in three sequential steps on manikin: Group A (step 1 blinded, no sensor), Group B (step 2 blinded sensor), Group C (step 3 sensor feedback). A weighing scale was placed below the manikin. This procedure was repeated once again at least 1 week apart. The feedback system used 3 different colours to indicate the force range achieved as below target, achieve target, above target. Significantly higher proportion of target cricoid pressure was achieved with the use of sensor feedback in Group C; 85.9% (95%CI: 82.7%-88.7%) compared to when blinded from sensor in Group B; 31.3% (95%CI: 27.4-35.4%). Cricoid force achieved blind (Group B) exceeded force achieved with feedback (Group C) by a mean of 8.0 (95%CI: 5.9-10.2, p |