Autor: |
Sieber Nicole, Lingenfelder Christian, Götz Madeleine, Heine Pauline, Lichtner Leonie, Hessling Martin |
Jazyk: |
angličtina |
Rok vydání: |
2021 |
Předmět: |
|
Zdroj: |
Current Directions in Biomedical Engineering, Vol 7, Iss 2, Pp 851-854 (2021) |
Druh dokumentu: |
article |
ISSN: |
2364-5504 |
DOI: |
10.1515/cdbme-2021-2217 |
Popis: |
Background: In current vitrectomy, the surgeon guides the vitrectome and a rigid fiber light guide with one hand each. It would be desirable to have a free hand for other surgical instruments and maneuvers. Methods: In the feasibility study presented here, a 20 gauge vitrectome is equipped with a miniature white LED that could eliminate the need for the separate light guide and therefore free one hand. The functionality of the system is proved in a ping-pong ball filled with agar and an ex-vivo porcine eye. Results: The brightness of the approach appears to be sufficient without any realistic phototoxic retinal hazard while the functionality of the vitrectome, in combination with the LED, is still given. Conclusion: The combination of a 20 gauge vitrectome with a LED illumination unit was successfully tested. The approach can be transferred to small vitrectomes in the future |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|