Comparison of intraocular pressure measurements between Icare PRO Tonometer, Goldmann Applanation Tonometer and non-contact tonometer in healthy and glaucomatous eyes
Autor: | Tatiana Sofia Monteiro Queirós, Márcia Clara Sá Rodrigues, Ana Carolina Pereira do Vale, Maria Josefina Costa Serino, Maria Teresa Leal Felgueiras Paínhas, Gabriel José Morgado Santos, Nuno Pedro Filipe da Costa e Lopes |
---|---|
Jazyk: | English<br />Spanish; Castilian<br />Portuguese |
Předmět: | |
Zdroj: | Revista Brasileira de Oftalmologia, Vol 77, Iss 5, Pp 248-254 |
Druh dokumentu: | article |
ISSN: | 1982-8551 0034-7280 |
DOI: | 10.5935/0034-7280.20180054 |
Popis: | Abstract Purpose: To validate the comparative measurements of intraocular pressure performed with IcarePRO tonometer (IPT) in relation to Goldmann applanation tonometer (GAT) and non-contact tonometer (NCT), as well as to evaluate the influence of central corneal thickness (CCT) on these values. Methods: This was a prospective and comparative study conducted during 6 months in the Department of Ophthalmology, Hospital de Braga, Portugal. The study population comprised two groups: healthy adults and adults with primary open angle glaucoma (POAG). The IOP values were obtained by the three tonometers and the order of execution was randomly selected. CCT values were obtained by ultrasonic pachymetry. Results: A total of 168 eyes (74 with diagnosis of POAG and 94 healthy) of 84 patients were included in this study. The mean IOP values obtained by IPT, GAT and NCT were 17.36±3.78 mmHg; 14.59±3.32 mmHg; and 17.04±4.01 mmHg, respectively. The comparison between IPT and NCT did not reveal statistically significant differences in the two groups studied. However, statistically significant differences were obtained between IPT and GAT values (p |
Databáze: | Directory of Open Access Journals |
Externí odkaz: |