Autor: |
Ramona C Dolscheid-Pommerich, Sarah Dolscheid, Lars Eichhorn, Birgit Stoffel-Wagner, Ingo Graeff |
Jazyk: |
angličtina |
Rok vydání: |
2018 |
Předmět: |
|
Zdroj: |
PLoS ONE, Vol 13, Iss 1, p e0190867 (2018) |
Druh dokumentu: |
article |
ISSN: |
1932-6203 |
DOI: |
10.1371/journal.pone.0190867 |
Popis: |
In acute stroke patients, thrombolysis is one gold standard therapy option within the first four hours after the ischemic event. A contraindication for thrombolysis is an International Normalized Ratio (INR) value >1.7. Since time is brain, rapid and reliable INR results are fundamental. Aim was to compare INR values determined by central laboratory (CL) analyzer and Point-of-Care Testing(POCT)-device and to evaluate the quality of POCT performance in cases of potential therapeutic thrombolysis at a certified stroke unit.In 153 patients INR measurements using POCT-devices (HEMOCHRON Signature Elite®) were compared to INR measurements (BCS®XP) performed at the central laboratory. Outlier evaluation was performed regarding the critical thrombolysis cut-off.Overall, we demonstrated a significant correlation (r = 0.809, p75 years.POCT-INR measurements based on our POCT concept are suitable to determine INR values in critical stroke patients. Nevertheless, outlier evaluation is mandatory. |