Influence Of Different Led Curing-Units On Depth Of Cure and Micro-Hardness Of Nano-Hybrid Resin Composite

Autor: Alper Kapdan, Digdem Eren, Seher Kaya
Jazyk: angličtina
Rok vydání: 2021
Předmět:
Zdroj: Cumhuriyet Dental Journal, Vol 24, Iss 1, Pp 88-95 (2021)
Druh dokumentu: article
ISSN: 1302-5805
2146-2852
DOI: 10.7126/cumudj.855065
Popis: Purpose: To evaluate the effect of two second generation LED Light curing units and a third-generation polywave LED Ligth curing unit at three different irradiation durations on surface microhardness and detpth of cure of nanofilled composite resin material. Materials and Methods: EliparTM S10, EliparTM Deep Cure-S, VALO Cordless, was evaluated at 10s, 20s, 40s curing periods in this study. A nanofilled composite FiltekTM Z550 that contains camphorquinone as photoinitiator is used as test material. For microhardness test ninety 6 mm diameter, 4 mm deep cylindrical composite blocks in teflon molds were prepared. For each light source at each application times 10 specimen were prepared (n=10) and tested. Also ninety, 4 mm diameter, 6 mm deep cylindrical composite specimens in a split aluminum mold polymerized with three different light sources at three different durations (n=10) were tested for depth of cure measurement. Data were analyzed statistically by SPSS (Ver: 22.0) and significance test of the difference between the two means (Kolmogorov-Simirnov), the variance analysis, and the Tukey Test used. Results: The values of cure depth were found significantly higher at 40 seconds of irradiation time for all light sources used (p