Reproducibility of goniometric measurement of the knee in the in-hospital phase following total knee arthroplasty

Autor: Geesink Ruud JT, Verhey Mark, Crijns Yvonne HF, van Dam Ellen M, Lenssen Anton F, van den Brandt Piet A, de Bie Rob A
Jazyk: angličtina
Rok vydání: 2007
Předmět:
Zdroj: BMC Musculoskeletal Disorders, Vol 8, Iss 1, p 83 (2007)
Druh dokumentu: article
ISSN: 1471-2474
DOI: 10.1186/1471-2474-8-83
Popis: Abstract Background The objective of the present study was to assess interobserver reproducibility (in terms of reliability and agreement) of active and passive measurements of knee RoM using a long arm goniometer, performed by trained physical therapists in a clinical setting in total knee arthroplasty patients, within the first four days after surgery. Methods Test-retest analysis Setting: University hospital departments of orthopaedics and physical therapy Participants: Two experienced physical therapists assessed 30 patients, three days after total knee arthroplasty. Main outcome measure: RoM measurement using a long-arm (50 cm) goniometer Agreement was calculated as the mean difference between observers ± 95% CI of this mean difference. The intraclass correlation coefficient (ICC) was calculated as a measure of reliability, based on two-way random effects analysis of variance. Results The lowest level of agreement was that for measurement of passive flexion with the patient in supine position (mean difference 1.4°; limits of agreement 16.2° to 19° for the difference between the two observers. The highest levels of agreement were found for measurement of passive flexion with the patient in sitting position and for measurement of passive extension (mean difference 2.7°; limits of agreement -6.7 to 12.1 and mean difference 2.2°; limits of agreement -6.2 to 10.6 degrees, respectively). The ability to differentiate between subjects ranged from 0.62 for measurement of passive extension to 0.89 for measurements of active flexion (ICC values). Conclusion Interobserver agreement for flexion as well as extension was only fair. When two different observers assess the same patients in the acute phase after total knee arthroplasty using a long arm goniometer, differences in RoM of less than eight degrees cannot be distinguished from measurement error. Reliability was found to be acceptable for comparison on group level, but poor for individual comparisons over time.
Databáze: Directory of Open Access Journals