Does standoff material affect acoustic radiation force impulse elastography? A preclinical study of a modified elastography phantom

Autor: Katharina Hollerieth, Bernhard Gaßmann, Stefan Wagenpfeil, Stephan Kemmner, Uwe Heemann, Konrad Friedrich Stock
Jazyk: angličtina
Rok vydání: 2018
Předmět:
Zdroj: Ultrasonography, Vol 37, Iss 2, Pp 140-148 (2018)
Druh dokumentu: article
ISSN: 2288-5919
2288-5943
DOI: 10.14366/usg.17002
Popis: Purpose This study was conducted to determine the influence of standoff material on acoustic radiation force impulse (ARFI) measurements in an elasticity phantom by using two different probes. Methods Using ARFI elastography, 10 observers measured the shear wave velocity (SWV, m/sec) in different lesions of an elasticity phantom with a convex 4C1 probe and a linear 9L4 probe. The experimental setup was expanded by the use of an interposed piece of porcine muscle as standoff material. The probe pressure on the phantom was registered. Results Faulty ARFI measurements occurred more often when quantifying the hardest lesion (74.0 kPa 4.97 m/sec) by the 9L4 probe with the porcine muscle as a standoff material interposed between the probe and the phantom. The success rate for ARFI measurements in these series was 52.4%, compared with 99.5% in the other series. The SWV values measured with the 9L4 probe were significantly higher (3.33±1.39 m/sec vs. 2.60±0.74 m/sec, P
Databáze: Directory of Open Access Journals