Comparison of Two Toric IOLs with Different Haptic Design: Optical Quality after 1 Year

Autor: Kata Miháltz, Michael Lasta, Michael Burgmüller, Pia Veronika Vécsei-Marlovits, Birgit Weingessel
Jazyk: angličtina
Rok vydání: 2018
Předmět:
Zdroj: Journal of Ophthalmology, Vol 2018 (2018)
Druh dokumentu: article
ISSN: 2090-004X
2090-0058
DOI: 10.1155/2018/4064369
Popis: Background. The purpose of this prospective, randomised study was to interocularly compare the visual performance after implantation of two different toric IOLs with different haptic design. Methods. 59 subjects with corneal astigmatism greater than 1.25 diopter (D) were implanted with an AT TORBI 709M IOL (Carl Zeiss Meditec AG) in one eye and with a Tecnis toric aspheric IOL (Abbot Medical Optics) in the other eye. Observation procedure was performed 12 months postoperatively. Main outcome measures included uncorrected distance visual acuity (UDVA), manifest refraction, IOL rotation, and IOL position. Results. Mean UCDVA was 0.04 ± 0.14 logMAR for AT TORBI eyes and 0.06 ± 0.15 logMAR for Tecnis eyes (p=0.3). The postoperative spherical equivalent values were significantly lower in the AT TORBI group. Mean toric IOL axis rotation was 3.0 ± 2.26 degrees for AT TORBI eyes and 3.27 ± 2.37 for Tecnis eyes (p=0.5). The mean vertical IOL tilt and vertical decentration values measured with the Visante OCT were significantly larger in the AT TORBI group (p
Databáze: Directory of Open Access Journals
Nepřihlášeným uživatelům se plný text nezobrazuje