Measuring Suite for Vascular Response Monitoring during Hyperbaric Oxygen Therapy by Means of Pulse Transit Time (PTT) Analysis

Autor: Theresa Wandel, Daniel Pascal Hausherr, Dirk Berben
Jazyk: angličtina
Rok vydání: 2022
Předmět:
Zdroj: Sensors, Vol 22, Iss 21, p 8295 (2022)
Druh dokumentu: article
ISSN: 1424-8220
DOI: 10.3390/s22218295
Popis: The efficacy of hyperbaric oxygen therapy in treating wound healing disorders is well established. The obvious explanation is the presence of elevated oxygen tissue tensions during the high-pressure oxygen exposure. This explanation omits that the effective agent, elevated oxygen tension, is only present for 6.25% of the time. To investigate possible prevailing vascular changes caused by HBOT, the presented device monitors the vascular response during therapy by Pulse-Transit-Time analysis. The device allows synchronous 1 kHz ECG and PPG measurements. The data are stored in a 1 GBit flash drive and retrieved post-therapy. Normoxic measurements on the authors with and without nicotine validate the device’s functionality. Measurements during HBO therapy have been successfully performed.
Databáze: Directory of Open Access Journals