Popis: |
Objectives: Therapeutic drug monitoring is strongly recommended for psychotropic drugs, which present a strong inter- and intra-individual variability due to multiple factors like inflammatory state, smoking, diet, drug interactions due to polypharmacy, and genetic profile. The aim of this study was to develop and validate a fast, simple, and sensitive method allowing the simultaneous quantification of a large number of psychotropic drugs. Methods: After a simple sample preparation with a one-step protein precipitation, a total of 55 compounds, including 22 antidepressants, 18 antipsychotics, 2 other psychotropic drugs (bupropion and nefopam), and their metabolites, was separated on a Waters Acquity HSS T3 ultra-performance liquid chromatography column, and subsequently detected and quantified by a triple quadrupole Quantis mass spectrometer with electrospray ionization operated in positive mode. Results: Total run time was only 5.7 min. Limits of detection ranged from 0.01 to 0.18 µg/L depending on compound. Measuring ranges were from 0.195 to 1000 µg/L depending on compound, and were defined according to therapeutic ranges. Inter- and intra-assay precisions values were less than 15 %. After validation, this method was successfully applied in daily practice for therapeutic drug monitoring of polymedicated psychiatric patients. Conclusion: We developed and validated one of the most sensitive and complete UPLC-MS/MS methods in psychopharmacology, allowing the simultaneous determination of 55 psychotropic drugs in only 5.7 min after a simple sample preparation. This method has been successfully used in daily practice for therapeutic drug monitoring of psychiatric patients and is especially useful in polymedicated patients. |