Localised Electrochemical Impedance Spectroscopy of Gold Nanoparticles Labelled Antibodies Probed by Platinum Microstructured Ultramicroelectrode
Autor: | Antanas Zinovicius, Inga Morkvenaite-Vilkonciene, Arunas Ramanavicius |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2024 |
Předmět: |
scanning electrochemical impedance microscopy
SEIM ultramicroelectrode electrode modification electrochemical immunosensor transducer Technology Electrical engineering. Electronics. Nuclear engineering TK1-9971 Engineering (General). Civil engineering (General) TA1-2040 Microscopy QH201-278.5 Descriptive and experimental mechanics QC120-168.85 |
Zdroj: | Materials, Vol 17, Iss 6, p 1339 (2024) |
Druh dokumentu: | article |
ISSN: | 1996-1944 |
DOI: | 10.3390/ma17061339 |
Popis: | This research is focused on enhancing the capabilities of scanning electrochemical impedance microscopy (SEIM) for detecting gold nanoparticle-labelled antibodies using electrochemically modified platinum ultramicroelectrode. The primary objective was to address the high resistance issue encountered in previous measurements with SEIM via the utilization of SEIM probes based on micro-electrodes modified by platinum microstructures, which improved the sensitivity and precision of the detection of targeted biomolecules. The modified probe resulted in a lowered charge transfer resistance by over ten times and a decrease in detection to around 100 fg/mL. We suggest potential applications in various biotechnological and biomedical fields, with future research expected to further refine this technique. |