Autor: |
Salma N. Ali, Samah S. Saad, Ahmed S. Fayed, Hoda M. Marzouk |
Jazyk: |
angličtina |
Rok vydání: |
2024 |
Předmět: |
|
Zdroj: |
BMC Chemistry, Vol 18, Iss 1, Pp 1-14 (2024) |
Druh dokumentu: |
article |
ISSN: |
2661-801X |
DOI: |
10.1186/s13065-024-01265-5 |
Popis: |
Abstract Ipratropium bromide (IPR) and fenoterol hydrobromide (FEN) have recently been combined in a promising inhaler to treat two prevalent inflammatory illnesses of the airways: bronchial asthma and chronic obstructive pulmonary disease (COPD). The necessity for a single, sensitive, and trustworthy analytical approach to cover the diverse and necessary tests of in-vitro and in-vivo studies is greatly grown with the rising production of new fixed combinations. Two novel, selective and environmentally friendly LC techniques were developed in order to guarantee precise measurement of IPR and FEN in their challenging formulation. The initial technique involved high-performance thin-layer chromatography (HPTLC) in conjunction with densitometric quantification. Chromatographic separation was attained on HPTLC plates utilizing ethyl acetate - ethanol - acetic acid (5.0:5.0:0.1, by volume) as a developing system. Densitometric quantification of the separated bands was carried out at 220.0 nm over concentration ranges of 0.50–15.0 µg/band for IPR and 0.50–12.0 µg/band for FEN. High-performance liquid chromatography (HPLC) paired with diode array detection (DAD) was the core of the second technique. The optimized separation was achieved on a Zorbax SB C18 (150 × 4.6 mm, 5 μm) column with a combination of 10.0 mM potassium dihydrogen orthophosphate, pH 5.0 ± 0.1, adjusted with o-phosphoric acid and methanol (70:30, v/v) as the mobile phase and pumped at flow rate of 1.0 mL/min. The peaks were monitored at 220.0 nm using diode array detection, achieving linearity range of 5.0–200.0 µg/mL for both drugs. The ICH criteria have been verified and both methods have been confirmed to be valid, and successfully applied for assay the cited drugs in the Atrovent® comp HFA metered dose inhaler as well as delivered dose uniformity testing of the final product. Finally, whiteness appraisal and several state-of-the-art green evaluation metrics were applied to evaluate the sustainability of the proposed methods. The suggested approaches produced promising results and are the first simple and sustainable methodologies for the simultaneous quantification of both drugs in different real samples, all of which strongly suggest their application in quality control laboratories. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|
Nepřihlášeným uživatelům se plný text nezobrazuje |
K zobrazení výsledku je třeba se přihlásit.
|