Model-based roentgen stereophotogrammetric analysis using elementary geometrical shape models: 10 years results of an uncemented acetabular cup component

Autor: Anne Jacobsen, Frank Seehaus, Yutong Hong, Han Cao, Alexander Schuh, Raimund Forst, Stefan Sesselmann
Jazyk: angličtina
Rok vydání: 2018
Předmět:
Zdroj: BMC Musculoskeletal Disorders, Vol 19, Iss 1, Pp 1-9 (2018)
Druh dokumentu: article
ISSN: 1471-2474
DOI: 10.1186/s12891-018-2259-4
Popis: Abstract Background Non-cemented acetabular cup components demonstrated different clinical performance depending on their surface texture or bearing couple. However, clinical osseointegration needs to be proved for each total joint arthroplasty (TJA) design. Aim of this study was to detect the in vivo migration pattern of a non-cemented cup design, using model-based roentgen stereophotogrammetric analysis with elementary geometrical shape models (EGS-RSA) to calculate early cup migration. Methods Interchangeable applicability of the model-based EGS-RSA method next to gold standard marker-based RSA method was assessed by clinical radiographs. Afterwards, in vivo acetabular cup migration for 39 patients in a maximum follow up of 120 months (10 years) was calculated using model-based EGS-RSA. Results For the axes with the best predictive capability for acetabular cup loosening, mean (±SD) values were calculated for migration and rotation of the cup. The cup migrated 0.16 (±0.22) mm along the cranio-caudal axis after 24 months and 0.36 (±0.72) mm after 120 months, respectively. It rotated − 0.61 (±0.57) deg. about the medio-lateral axis after 24 months and − 0.53 (±0.67) deg. after 120 months, respectively. Conclusions Interchangeable applicability of model-based EGS-RSA next to gold standard marker-based RSA method could be shown. Model-based EGS-RSA enables an in vivo migration measurement without the necessity of TJA specific surface models. Migration of the investigated acetabular cup component indicates significant migration values along all the three axes. However, migration values after the second postoperative year were within the thresholds reported in literature, indicating no risk for later aseptic component loosening of this TJA design.
Databáze: Directory of Open Access Journals
Nepřihlášeným uživatelům se plný text nezobrazuje