Popis: |
Abstract Background Previous studies have confirmed the feasibility of the cortical bone trajectory (CBT) technique. However, there are few reports on spinous process violation and screw penetration during the screw insertion. The purpose of this study was to evaluate the incidence of spinous process violation and screw penetration through the pedicle during CBT screw insertion. Methods Computed tomography (CT) scans with normal lumbar structures were consecutively obtained and three-dimensional (3D) reconstructions of the lumbar spine were created. Bilateral CBT screw placement was simulated on each segment using a screw diameter of 4.5 mm, 5.0 mm, or 5.5 mm. Incidences of these complications were recorded and analyzed. Results A total of 90 patients were enrolled. Spinous process violation was observed in 68.3, 53.3, 25.5, 1.7, and 0% from L1 to L5, respectively, using 4.5 mm screws. A significant difference was found among the five segments but this was unconnected to gender or screw diameter. The incidence of screw penetration through the inner wall decreased from L1 to L4; in turn, L1 (16.7–35.5%), L2 (12.7–34.4%), L3 (2.8–23.8%) and L4 (1.1–6.7%). This trend was reversed in L5 (6.7–16.7%). Moreover, screw penetration through the outer wall was rare. The incidence of screw penetration varied with screw size as well as lumbar level, but not with gender. Conclusions There are more difficulties of CBT screw fixation in upper lumbar spine. The low rate of screw penetration, using 4.5 mm screws, suggests the safety for CBT fixation in the lumbar spine. Larger screws (5.0 mm or 5.5 mm) are more recommended for use in the lower lumbar spine. Moreover, CBT fixation in L5 deserves greater attention because of the unique morphology of the pedicle. |