Release of titanium after insertion of dental implants with different surface characteristics – an ex vivo animal study

Autor: Mattias Pettersson, Jean Pettersson, Margareta Molin Thorén, Anders Johansson
Jazyk: angličtina
Rok vydání: 2017
Předmět:
Zdroj: Acta Biomaterialia Odontologica Scandinavica, Vol 3, Iss 1, Pp 63-73 (2017)
Druh dokumentu: article
ISSN: 2333-7931
23337931
DOI: 10.1080/23337931.2017.1399270
Popis: In the present study, amount of titanium (Ti) released into the surrounding bone during placement of implants with different surface structure was investigated. Quantification of Ti released during insertion from three different implants was performed in this ex vivo study. Jaw bone from pigs was used as model for installation of the implants and Inductively Coupled Plasma Atomic Emission Spectroscopy (ICP-AES) was used for analysis of the released Ti. Implant surface were examined with scanning electron microscopy (SEM), before and after the placement into the bone. Ti was abraded to the surrounding bone upon insertion of a dental implant and the surface roughness of the implant increased the amount of Ti found. Diameter and total area of the implant were of less importance for the Ti released to the bone. No visible damages to the implant surfaces could be identified in SEM after placement.