Platelet count in impedance-based hematology analyzer: Beware of trap!
Autor: | Vani Krishnamurthy, Archana Shivamurthy, Pradeep V Kumar |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2023 |
Předmět: | |
Zdroj: | Asian Journal of Transfusion Science, Vol 17, Iss 1, Pp 131-132 (2023) |
Druh dokumentu: | article |
ISSN: | 0973-6247 1998-3565 |
DOI: | 10.4103/ajts.ajts_65_21 |
Popis: | Among the blood counts, platelet count is most often reported with inconsistency. Many of the analyzers work on electrical impedance principle for red blood cell (RBC) and platelet counting. However, with this technology, factors such as fragmented RBCs, microcytes, cytoplasmic fragments of leukemic cells, lipid particles, fungal yeast forms, and bacteria are known to interfere with platelet count and give spuriously elevated platelet counts. A 72-year-old male was admitted for the treatment of dengue infection who had serial platelet count monitoring. He had an initial platelet count of 48,000/cumm which suddenly improved to 2.6 lakhs within 6 h without any platelet transfusion. Peripheral smear however did not correlate with the machine-derived count. Repeat test after 6 h yielded a result of 56,000/cumm which correlated well with the peripheral smear. This falsely elevated count was due to the presence of lipid particles as the sample was drawn in the postprandial state. |