Autor: |
Stefan Krennmair, Michael Malek, Raphael Stehrer, Philip Stähler, Sven Otto, Lukas Postl |
Jazyk: |
angličtina |
Rok vydání: |
2023 |
Předmět: |
|
Zdroj: |
European Journal of Medical Research, Vol 28, Iss 1, Pp 1-11 (2023) |
Druh dokumentu: |
article |
ISSN: |
2047-783X |
DOI: |
10.1186/s40001-023-01580-y |
Popis: |
Abstract Purpose The present three-dimensional (3D) finite element analysis (FEA) was aimed to assess the biomechanical effects and fracture risks of four different interforaminal implant-prosthodontic anchoring configurations exposed to frontal trauma. Material and methods A symphyseal frontal trauma of 1 MPa was applied to four dental implant models with different configurations (two unsplinted interforaminal implants [2IF-U], two splinted interforaminal implants [2IF-S], four unsplinted interforaminal implants[ 4IF-U], four splinted interforaminal implants [4IF-S]. By using a 3D-FEA analysis the effective cortical bone stress values were evaluated in four defined regions of interest (ROI) (ROI 1: symphyseal area; ROI 2: preforaminal area; ROI 3: mental foraminal area; and ROI 4: condylar neck) followed by a subsequent intermodel comparison. Results In all models the frontal traumatic force application revealed the highest stress values in the condylar neck region. In both models with a four-implant configuration (4IF-U, 4IF-S), the stress values in the median mandibular body (ROI 1) and in the condylar neck region (ROI 4) were significantly reduced (P |