A multicenter study comparing the bacterial reduction on flexible endoscopes without a working channel between UV-C light disinfection versus standard endoscope Washer Disinfection: a randomized controlled trial

Autor: Yana Halmans, D. J. Wellenstein, M. Romijn, A. J. M. van Bemmel, H. van den Berge, R. A. Scheeren, J. S. Kalpoe, R. Klont, J. H. van Zeijl, H. Sikkema, S. M. Euser, J. Hopman, R. P. Takes, G. B. van den Broek
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: Antimicrobial Resistance and Infection Control, Vol 13, Iss 1, Pp 1-8 (2024)
Druh dokumentu: article
ISSN: 2047-2994
DOI: 10.1186/s13756-024-01486-2
Popis: Abstract Background To prevent cross-contamination between patients, adequate reprocessing is necessary when using flexible endoscopes (FEs) without a working channel. The current reprocessing process using an Endoscope Washer Disinfector (EWD) is time-consuming. Ultraviolet light group C (UV-C) exposition is an alternative and fast disinfection method and has previously been shown to adequately reduce Colony Forming Units (CFUs) on FEs without a working channel. The objective of this study was to examine whether UV-C light is as effective in reducing CFUs on contaminated FEs without a working channel compared to the EWD. Methods FEs without a working channel were collected in three different Otorhinolaryngology Departments in the Netherlands. After pharyngolaryngoscopy, a manual pre-cleaning with tap water was performed and a culture was collected by rolling the distal 8–10 cm of the FE over an agar plate. Next, the FE was randomly assigned to be disinfected with UV-C light (D60) or the EWD (gold standard). After disinfection, another culture was taken. The primary outcome was microbiological contamination, defined by Colony Forming Units (CFU). Results 600 FEs without a working channel were randomized. After clinical use and manual pre-cleaning, 239/300 (79.7%) FEs in the UV-C group and 262/300 (87.3%) FEs in the EWD group were contaminated (i.e., > 0 CFU). FEs without culture confirmed contamination were excluded from further analysis. After UV-C light disinfection, 195/239 (81.6%) FEs showed 0 CFUs, compared to 187/262 (71.4%) FEs disinfected with the EWD (p
Databáze: Directory of Open Access Journals
Nepřihlášeným uživatelům se plný text nezobrazuje