Electrodermal Activity Sensor for Classification of Calm/Distress Condition

Autor: Roberto Zangróniz, Arturo Martínez-Rodrigo, José Manuel Pastor, María T. López, Antonio Fernández-Caballero
Jazyk: angličtina
Rok vydání: 2017
Předmět:
Zdroj: Sensors, Vol 17, Iss 10, p 2324 (2017)
Druh dokumentu: article
ISSN: 1424-8220
17102324
DOI: 10.3390/s17102324
Popis: This article introduces a new and unobtrusive wearable monitoring device based on electrodermal activity (EDA) to be used in health-related computing systems. This paper introduces the description of the wearable device capable of acquiring the EDA of a subject in order to detect his/her calm/distress condition from the acquired physiological signals. The lightweight wearable device is placed in the wrist of the subject to allow continuous physiological measurements. With the aim of validating the correct operation of the wearable EDA device, pictures from the International Affective Picture System are used in a control experiment involving fifty participants. The collected signals are processed, features are extracted and a statistical analysis is performed on the calm/distress condition classification. The results show that the wearable device solely based on EDA signal processing reports around 89% accuracy when distinguishing calm condition from distress condition.
Databáze: Directory of Open Access Journals