Autor: |
Hairenanorashikin Sharip, Rafidah Supar |
Jazyk: |
angličtina |
Rok vydání: |
2020 |
Předmět: |
|
Zdroj: |
Journal of Academia, Vol 8, Iss 1, Pp 18-23 (2020) |
Druh dokumentu: |
article |
ISSN: |
2289-6368 |
Popis: |
This experimental study investigated the significant different in image quality and dose between different automatic exposure of exposure parameter (AOP) mode in full field digital mammography (FFDM) system. CIRS012A and PMMA breast phantom (4cm, 5cm and 6 cm thickness) were used as subject using Senographe Essential FFDM system. TLD chip was used as dose measurement tool. Exposures were taken in cranio-caudal projection. 2 radiographers with more than 10 years of working experience in performing mammography scored the image independently. Kappa finds a good agreement between raters (kappa value=0.9, p0.05). DOSE mode is the preferred selection in optimizing between dose and image quality. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|