Popis: |
Purpose. To compare two threshold strategies for visual field assessment, ZETA Fast (Optopol Technology) and Humphrey SITA Fast (Carl Zeiss Meditec), in controls and subjects with glaucoma. Patients and Methods. A prospective case-control study was carried out in which the clinical practice study included 26 controls and 26 glaucoma subjects. Testing for each strategy was monocular. Quantitative comparisons of mean deviation (MD), pattern standard deviation (PSD), visual field index (VFI), and test duration were made using two one-sided t-tests and Wilcoxon signed-rank tests. Confusion matrices were constructed to assess Optopol’s detection as a proxy for Zeiss’s detection of early glaucomatous defects. Receiver operating characteristic (ROC) curves were used to assess MD and PSD’s discriminability. Results. The difference in MD values (Optopol-Zeiss) was within the margin for controls (difference = 0.36, p=0.06), but not for glaucomatous subjects (difference = 2.16, p=1.0). The Optopol strategy took longer than the Zeiss strategy in both controls (difference = 23 seconds, p=0.001) and glaucomatous subjects (difference = 49 seconds, p |