Comparison of the effectiveness of different high-flow devices in neonatal care

Autor: Diogo Machado Kaminski, Graziela Ferreira Biazus, Rita C. Silveira, Renato Soibelmann Procianoy
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: Jornal de Pediatria, Vol 100, Iss 2, Pp 212-217 (2024)
Druh dokumentu: article
ISSN: 0021-7557
DOI: 10.1016/j.jped.2023.10.006
Popis: Objective: To evaluate the success rate of high-flow nasal cannula (HFNC) therapy using an adapted obsolete mechanical ventilator (MV), Optiflow™ and Vapotherm™ in newborns (NBs). Method: This was a retrospective observational study conducted in the neonatal intensive care unit (NICU). The sample comprised NBs who underwent HFNC therapy due to ventilatory dysfunction, for weaning from non-invasive ventilation (NIV), or post-extubation. The three groups, stratified according to gestational age (GA) and birth weight, and corrected GA and weight at the beginning of HFNC use, were as follows: Optiflow ™, Vapotherm ™, and obsolete Mechanical Ventilator (MV) adapted for high flow therapy. Subsequently, the NBs were divided into a success group (SG) and a failure group (FG). HFNC success was defined as a therapy duration exceeding 72 h. Results: A total of 209 NBs were evaluated, with 31.1 % using HFNC due to ventilatory dysfunction, 2.4 % after extubation, and 66.5 % after NIV weaning. HFNC success rate was observed in 90.9 % of the NBs, with no difference between equipment types (Vapotherm ™, Optiflow ™, and adapted VM). Conclusion: Different types of HFNC equipment are equally effective when used in neonatology for respiratory dysfunction, as a method of weaning from NIV and post-extubation. Adapted obsolete MV can be an alternative for HFCN therapy in resource-constrained settings.
Databáze: Directory of Open Access Journals