Adverse Soft-Tissue Reaction After Ceramic-On-Ceramic Bearing Fracture Mimicking a Periprosthetic Joint Infection

Autor: Ruben Tresgallo-Parés, MD, Norberto J. Torres-Lugo, MD, Edwin Rosado-Hernández, MD, Gerardo Olivella, MD, MPH, Norman Ramírez, MD, Antonio Otero-López, MD
Jazyk: angličtina
Rok vydání: 2022
Předmět:
Zdroj: Arthroplasty Today, Vol 13, Iss , Pp 93-97 (2022)
Druh dokumentu: article
ISSN: 2352-3441
DOI: 10.1016/j.artd.2021.12.001
Popis: Prosthetic failure secondary to bearing fracture remains a potential complication in ceramic-on-ceramic total hip arthroplasty. We report the unusual presentation of a ceramic component fracture of a total hip arthroplasty performed 17 years ago that mimicked a periprosthetic joint infection. This case was managed based on the current guidelines and algorithms recommended for periprosthetic joint infection management. Histologic examination of periprosthetic tissue revealed an adverse inflammatory soft-tissue reaction to the ceramic fragments released from the fracture site. Our case highlights a misleading, inflammatory acute response usually associated with an infectious process corresponding to an adverse soft-tissue reaction. High clinical suspicion and a systematic approach are essential to address these deceiving clinical scenarios.