Autor: |
Surapon Tangvarasittichai, Prawit Taytiwat |
Jazyk: |
angličtina |
Rok vydání: |
2020 |
Předmět: |
|
Zdroj: |
Siriraj Medical Journal, Vol 53, Iss 7 (2020) |
Druh dokumentu: |
article |
ISSN: |
2228-8082 |
Popis: |
High density lipoprotein (HDL) are responsible for the reverse transport of cholesterol from the peripheral cells to the liver. Here, cholesterol is metabolised. Monitoring of HDL-C in serum is of clinical importance since an inverse correlation exists between serum HDL-C concentrations and the risk of atherosclerotic disease. Elevated HDL-C concentrations are protective against coronary heart disease. In this study, an automated enzymatic method for the direct determination of HDL-C in serum was found to have good precision and accuracy. Reproducibility was determined daily using control Precinorm L® in an internal protocol (n=20), within-run CV = 1.3% and between-day CV = 2.013%. The comparison of the HDL-C enzymatic assay with HDL-C precipitating agent was good with a correlation coefficient r = 0.965, Y = 1.981+ 0.959X, n =172. The normal range of HDL-C concentration in serum by this enzymatic method was 36.16 -86.08 mg/dl. Thus, a HDL-C level of over 35 mg/dl, shows the risk is not high and is protective against coronary heart disease. |