Effect of individualized PEEP titration guided by intratidal compliance profile analysis on regional ventilation assessed by electrical impedance tomography – a randomized controlled trial

Autor: Jonas Weber, Jan Gutjahr, Johannes Schmidt, Sara Lozano-Zahonero, Silke Borgmann, Stefan Schumann, Steffen Wirth
Jazyk: angličtina
Rok vydání: 2020
Předmět:
Zdroj: BMC Anesthesiology, Vol 20, Iss 1, Pp 1-10 (2020)
Druh dokumentu: article
ISSN: 1471-2253
DOI: 10.1186/s12871-020-00960-9
Popis: Abstract Background The application of positive end-expiratory pressure (PEEP) may reduce dynamic strain during mechanical ventilation. Although numerous approaches for PEEP titration have been proposed, there is no accepted strategy for titrating optimal PEEP. By analyzing intratidal compliance profiles, PEEP may be individually titrated for patients. Methods After obtaining informed consent, 60 consecutive patients undergoing general anesthesia were randomly allocated to mechanical ventilation with PEEP 5 cmH2O (control group) or PEEP individually titrated, guided by an analysis of the intratidal compliance profile (intervention group). The primary endpoint was the frequency of each nonlinear intratidal compliance (CRS) profile of the respiratory system (horizontal, increasing, decreasing, and mixed). The secondary endpoints measured were respiratory mechanics, hemodynamic variables, and regional ventilation, which was assessed via electrical impedance tomography. Results The frequencies of the CRS profiles were comparable between the groups. Besides PEEP [control: 5.0 (0.0), intervention: 5.8 (1.1) cmH2O, p