Autor: |
Michael Martin Eder, Michael Reiner, Christian Heinz, Sylvia Garny, Philipp Freislederer, Guillaume Landry, Maximilian Niyazi, Claus Belka, Marco Riboldi |
Jazyk: |
angličtina |
Rok vydání: |
2022 |
Předmět: |
|
Zdroj: |
Zeitschrift für Medizinische Physik, Vol 32, Iss 3, Pp 296-311 (2022) |
Druh dokumentu: |
article |
ISSN: |
0939-3889 |
DOI: |
10.1016/j.zemedi.2022.02.005 |
Popis: |
Frameless single-isocenter non-coplanar stereotactic radiosurgery (SRS) for patients with multiple brain metastases is a treatment at high geometrical complexity. The goal of this study is to analyze the dosimetric impact of non-coplanar image guidance with stereoscopic X-ray imaging. Such an analysis is meant to provide insights on the adequacy of safety margins, and to evaluate the benefit of imaging at non-coplanar configurations.The ExacTrac® (ET) system (Brainlab AG, Munich, Germany) was used for stereoscopic X-ray imaging in frameless single-isocenter non-coplanar SRS for multiple brain metastases. Sub-millimeter precision was found for the ET-based pre-treatment setup, whereas a degradation was noted for non-coplanar treatment angles. Misalignments without intra-fractional positioning corrections were reconstructed in 6 degrees of freedom (DoF) to resemble the situation without non-coplanar image guidance.Dose recalculation in 20 SRS patients with applied positioning corrections did not reveal any significant differences in D98% for 75 planning target volumes (PTVs) and gross tumor volumes (GTVs). For recalculation without applied positioning corrections, significant differences (p |