Autor: |
Giorgia Spinazzola, Giuliano Ferrone, Roberta Costa, Marco Piastra, Gianmarco Maresca, Marco Rossi, Massimo Antonelli, Giorgio Conti |
Jazyk: |
angličtina |
Rok vydání: |
2023 |
Předmět: |
|
Zdroj: |
BMC Pulmonary Medicine, Vol 23, Iss 1, Pp 1-10 (2023) |
Druh dokumentu: |
article |
ISSN: |
1471-2466 |
DOI: |
10.1186/s12890-023-02489-2 |
Popis: |
Abstract Historically, the oro-nasal mask has been the preferred interface to deliver Non-Invasive Positive Pressure Ventilation (NPPV) in critically ill patients. To overcome the problems related to air leaks and discomfort, Total Full-face masks have been designed. No study has comparatively evaluated the performance of the total Full-face masks available.The aim of this bench study was to evaluate the influence of three largely diffuse models of total Full -face masks on patient-ventilator synchrony and performance during pressure support ventilation. NPPV was applied to a mannequin, connected to an active test lung through three largely diffuse Full-face masks: Dimar Full-face mask (DFFM), Performax Full-face mask (RFFM) and Pulmodyne Full-face mask (PFFM).The performance analysis showed that the ΔPtrigger was significantly lower with PFFM (p |