Dynamics of manual impaction instruments during total hip arthroplasty

Autor: Adam Reynolds, Ruben Doyle, Oliver Boughton, Justin Cobb, Sarah Muirhead-Allwood, Jonathan Jeffers
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: Bone & Joint Research, Vol 13, Iss 4, Pp 193-200 (2024)
Druh dokumentu: article
ISSN: 2046-3758
DOI: 10.1302/2046-3758.134.BJR-2023-0224.R1
Popis: Aims: Manual impaction, with a mallet and introducer, remains the standard method of installing cementless acetabular cups during total hip arthroplasty (THA). This study aims to quantify the accuracy and precision of manual impaction strikes during the seating of an acetabular component. This understanding aims to help improve impaction surgical techniques and inform the development of future technologies. Methods: Posterior approach THAs were carried out on three cadavers by an expert orthopaedic surgeon. An instrumented mallet and introducer were used to insert cementless acetabular cups. The motion of the mallet, relative to the introducer, was analyzed for a total of 110 strikes split into low-, medium-, and high-effort strikes. Three parameters were extracted from these data: strike vector, strike offset, and mallet face alignment. Results: The force vector of the mallet strike, relative to the introducer axis, was misaligned by an average of 18.1°, resulting in an average wasted strike energy of 6.1%. Furthermore, the mean strike offset was 19.8 mm from the centre of the introducer axis and the mallet face, relative to the introducer strike face, was misaligned by a mean angle of 15.2° from the introducer strike face. Conclusion: The direction of the impact vector in manual impaction lacks both accuracy and precision. There is an opportunity to improve this through more advanced impaction instruments or surgical training. Cite this article: Bone Joint Res 2024;13(4):193–200.
Databáze: Directory of Open Access Journals