Autor: |
Rongshan Li, Ruixue Ran, Quansheng Li, Yurong Huang, Yuan Gu, Duanyun Si |
Jazyk: |
angličtina |
Rok vydání: |
2016 |
Předmět: |
|
Zdroj: |
Journal of Pharmaceutical Analysis, Vol 6, Iss 3, Pp 184-189 (2016) |
Druh dokumentu: |
article |
ISSN: |
2095-1779 |
DOI: |
10.1016/j.jpha.2016.03.002 |
Popis: |
Deoxyglycychloxazol (TY501) is a glycyrrhetinic acid derivative which exhibits high anti-inflammatory activity and reduced pseudoaldosteronism compared to glycyrrhetinic acid. In this study, a sensitive and rapid liquid chromatography–tandem mass spectrometry (LC–MS/MS) method was established for the quantitation of TY501 in rat plasma. Plasma samples were treated by precipitating protein with methanol and supernatants were separated by a Symmetry C8 column with the mobile phase consisting of methanol and 10 mM ammonium formate (containing 0.1% of formic acid) (90:10, v/v). The selected reaction monitoring (SRM) transitions were performed at m/z 647.4→191.2 for TY501 and m/z 473.3→143.3 for astragaloside aglycone (IS) in the positive ion mode with atmospheric pressure chemical ionization (APCI) source. Calibration curve was linear over the concentration range of 5–5000 ng/mL. The lower limit of quantification was 5 ng/mL. The mean recovery was over 88%. The intra- and inter-day precisions were lower than 6.0% and 12.8%, respectively, and the accuracy was within ±1.3%. TY501 was stable under usual storage conditions and handling procedure. The validated method has been successfully applied to a pharmacokinetic study after oral administration of TY501 to rats at a dosage of 10 mg/kg. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|