Comparison of the Chemiluminescence Immunoassay LIAISON® with the Radioimmunoassay for Aldosterone and Renin Measurement
Autor: | Jana Uhrová, Hana Benáková, Zdislava Vaníčková, Tomáš Zima |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2021 |
Předmět: | |
Zdroj: | Prague Medical Report, Vol 122, Iss 2, Pp 80-95 (2021) |
Druh dokumentu: | article |
ISSN: | 1214-6994 2336-2936 23362936 |
DOI: | 10.14712/23362936.2021.9 |
Popis: | Determination of renin plasma levels is useful in the diagnosis of hypertension and in the therapeutic follow-up of hypertensive patients. Plasmatic concentration of renin decreases in patients with hypertension due to a primary hyperaldosteronism, contrary to renovascular hypertension where concentrations of renin and aldosterone are both elevated. Blood samples (serum, EDTA plasma) were analysed using two different chemiluminiscent methods CLIA LIAISON® and radioimmunoassay for aldosterone (IMMUNOTECH Beckman Coulter) and renin (Cisbio Bioassay) measurements were compared. We used both methods to ascertain the correlation between serum vs. EDTA plasma levels of aldosterone (RIA, CLIA) and renin (IRMA, CLIA) and to compare aldosterone to renin ratios for CLIA and for radioimmunoassay: serum aldosterone to plasma renin and plasma aldosterone to plasma renin. We compared serum aldosterone CLIA vs. RIA (rP=0.933, P |