Design of reduced size long counter with flat response function from few keV to 20 MeV

Autor: Mazunga Mohamed, JI Xiang, Li Taosheng, Li Yanan
Jazyk: angličtina
Rok vydání: 2016
Předmět:
Zdroj: Nuclear Technology and Radiation Protection, Vol 31, Iss 4, Pp 361-365 (2016)
Druh dokumentu: article
ISSN: 1451-3994
1452-8185
DOI: 10.2298/NTRP1604361M
Popis: In this study, a reduced size neutron long counter with improved response function was designed using SuperMC simulation code. The main parts of the reduced size long counter are 3He proportional counter, cylindrical moderators made of polyethylene, chromium metal converter, neutron absorber material and outer aluminum cover. A 1.5 cm thick layer of chromium metal was embedded into the hydrogen-rich moderator to enhance response of the long counter at high neutron energies. The radius and length of the long counter was reduced to 15 cm and 35 cm, respectively. In this design we managed to obtain a flat response function from a few keV to 20 MeV energy range. The angular response function, determined at 0.5 MeV and 14 MeV neutron energies, confirmed that the reduced size long counter designed in this study is not significantly affected by in-scattered neutrons. The reduced size long counter is suitable for use as a standard transfer instrument for monitoring neutron fluence in mono-energetic neutron fields.
Databáze: Directory of Open Access Journals