Establishing and Verifying a Robust Liquid Chromatography-Tandem Mass Spectrometry Method to Simultaneously Measure Seven Androgens Present in Plasma Samples

Autor: Songlin Yu, Yutong Zou, Yicong Yin, Jialei Yu, Qianqian Li, Shaowei Xie, Wei Luo, Xiaoli Ma, Danchen Wang, Ling Qiu
Jazyk: angličtina
Rok vydání: 2022
Předmět:
Zdroj: Separations, Vol 9, Iss 11, p 377 (2022)
Druh dokumentu: article
ISSN: 2297-8739
DOI: 10.3390/separations9110377
Popis: Objectives: To develop a robust liquid chromatography-tandem mass spectrometry (LC–MS/MS) method to simultaneously measure seven human plasma androgens, namely testosterone (T), dihydrotestosterone (DHT), androstenedione (A4), dehydroepiandrosterone sulfate (DHEAS), dehydroepiandrosterone (DHEA), 11-ketotestosterone (11-KetoT), and 11β-hydroxytestosterone (11β-OHT). Design and Methods: Plasma was extracted via a solid phase extraction method, and the analytical performance of the assay was validated according to the Clinical & Laboratory Standards Institute guidelines. Overall, 73 apparently healthy volunteers were recruited to evaluate the distribution of these seven androgens; their levels in 25 females with acne and 33 obese females were also evaluated. Results: The developed method exhibited a good precision, with the total coefficient variations (CV) and the intra-assay CVs being within 10%. Furthermore, the recoveries of T, DHT, A4, DHEA, DHEAS, 11-KetoT, and 11β-OHT were 90.3–105.8, 88.7–98.1, 92.4–102.5, 90.5–106.7, 87.6–99.9, 93.3–105.3, and 90.2–104.4%, respectively, and no significant matrix effect was observed after internal standard correction (
Databáze: Directory of Open Access Journals