Autor: |
Yuanshan Liu, Tianen Zhou, Qiyu Yang, Yujing Lu, Zhengfei Yang, Jun Jiang |
Jazyk: |
angličtina |
Rok vydání: |
2024 |
Předmět: |
|
Zdroj: |
Resuscitation Plus, Vol 17, Iss , Pp 100572- (2024) |
Druh dokumentu: |
article |
ISSN: |
2666-5204 |
DOI: |
10.1016/j.resplu.2024.100572 |
Popis: |
Aim: Spectral Flux (SF), which is based on common algorithms in the audio processing field, was applied to quantitatively analyze ECG signals to optimize the timing of defibrillation. With the aim of proving the performance in optimizing the timing of defibrillation, SF was compared with Amplitude Spectrum Area (AMSA) in a porcine model of ventricular fibrillation (VF) in a retrospective analysis experiment. Methods: A total of 56 male domestic pigs, weighing 40 ± 5 kg, were induced to undergo VF. Animals were then left untreated for 10 min, and after 6 min of cardiopulmonary resuscitation (CPR) defibrillation was performed. The respective SF and AMSA values were calculated every minute during VF and CPR. Comparisons were made through receiver operating characteristic (ROC) curves, one-way analyses of variance (one-way ANOVA), and scatterplots for the successful initial defibrillation sample (positive samples, Group R) and the failed initial defibrillation sample (negative samples, Group N) to illustrate the performance in optimizing the timing of defibrillation for the AMSA and SF methods. Result: Values of SF and AMSA gradually decreased during the 10 min VF period and increased in during the 6 min CPR period. The scatterplots showed that both metrics had the ability to distinguish positive and negative samples (p |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|