Autor: |
Fabienne Dobbels, Johan Vanhaecke, Sofie Vandenbroeck, Sabina De Geest, Leentje De Bleser |
Jazyk: |
angličtina |
Rok vydání: |
2010 |
Předmět: |
|
Zdroj: |
Sensors, Vol 10, Iss 3, Pp 1652-1660 (2010) |
Druh dokumentu: |
article |
ISSN: |
1424-8220 |
DOI: |
10.3390/s100301652 |
Popis: |
In a prospective descriptive laboratory study, 25 Helping Hand™ (HH) (10 without and 15 with reminder system) and 50 Medication Event Monitoring Systems (MEMS) (25 with 18-month and 25 with 2-year battery life) were manipulated twice daily following a predefined protocol during 3 consecutive weeks. Accuracy was determined using the fixed manipulation scheme as the reference. Perfect functioning (i.e., total absence of missing registrations and/or overregistrations) was observed in 70% of the HH without, 87% of the HH with reminder, 20% MEMS with 18 months, and 100% with 2-year battery life respectively. |
Databáze: |
Directory of Open Access Journals |