Simulation of correction factor about electronic loss and fluorescence and scattering for W/MO X-rays
Autor: | Ji Wang, Shiwei Ren, Jinjie Wu, Siming Guo, Hongjie Li |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2019 |
Předmět: |
ionisation chambers
dosimetry tungsten molybdenum nuclear electronics x-ray detection voltage 23.0 kv to 35.0 kv ionisation chamber w/mo x-rays tungsten target scattering correction factors electronic loss correction factor low effective energy molybdenum-filtered x-ray radiation fluorescence scattering correction factor electron loss correction factor low-energy free air low-energy radiation field monte–carlo program egsnrc Engineering (General). Civil engineering (General) TA1-2040 |
Zdroj: | The Journal of Engineering (2019) |
Druh dokumentu: | article |
ISSN: | 2051-3305 |
DOI: | 10.1049/joe.2018.9097 |
Popis: | The Monte–Carlo (MC) program EGSnrc was used to establish the model of low-energy radiation field and low-energy free air ionisation chamber. The electron loss correction factor and the fluorescence scattering correction factor of the molybdenum-filtered X-ray radiation of 23 ∼ 35kv tungsten target were simulated. The results show that due to the low effective energy of the radiation and the energy is close, the influence of the electronic loss correction factor is negligible, and the fluorescence and scattering correction factors increase slightly with the increase of the effective energy. |
Databáze: | Directory of Open Access Journals |
Externí odkaz: |