Autor: |
Ma. Stella Grace B. Lequiron, Daniel J. Joe, Sun Do Lim, Il Doh |
Jazyk: |
angličtina |
Rok vydání: |
2023 |
Předmět: |
|
Zdroj: |
Applied Sciences, Vol 13, Iss 10, p 6220 (2023) |
Druh dokumentu: |
article |
ISSN: |
2076-3417 |
DOI: |
10.3390/app13106220 |
Popis: |
The thermal response time of a thin-film resistance temperature detector (RTD) array sensor was measured for a high-intensity focused ultrasound (HIFU) phantom. As the temperature inside materials change rapidly within several seconds, it is important to have a temperature sensor with a fast response time to evaluate their performance. However, previous methods for measuring thermal response time were not suitable for thin-film sensors, and there were no quantitative data available. In this study, we used a liquid drop method to measure the thermal time constant of the thin-film RTD, which was found to be 1.0 ± 0.2 ms. This indicates that the thin-film RTD array sensor has a sufficiently fast response time to detect sudden temperature changes inside the tissue-mimicking material (TMM) for validating HIFU devices. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|