Diagnostic Accuracy of Charge-coupled Device Sensor and Photostimulable Phosphor Plate Receptor in the Detection of External Root Resorption In Vitro
Autor: | Shirin Sakhdari, Zohreh Khalilak, Elham Najafi, Rezvaneh Cheraghi |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2015 |
Předmět: | |
Zdroj: | Journal of Dental Research, Dental Clinics, Dental Prospects, Vol 9, Iss 1, Pp 18-22 (2015) |
Druh dokumentu: | article |
ISSN: | 2008-210X 2008-2118 |
DOI: | 10.15171/joddd.2015.004 |
Popis: | Background and aims. Early diagnosis of external root resorption is important for accurate treatment. The purpose of this study was to compare the efficacy of a charge-coupled device (CCD) sensor and a photostimulable phosphor (PSP) plate receptor in the diagnosis of artificial external root resorption. Materials and methods. In this diagnostic in-vitro study, 40 maxillary incisors were mounted in a segment of dry bone and preliminary radiographs were obtained using CCD and PSP sensors. Artificial resorption cavities were produced on the middle-third in half of the samples and on the cervical-third in the other half on the buccal root surfaces. Radiographs were repeated and images were evaluated. Data were statistically analyzed using chi-square and diagnostic tests. Results. There were no significant differences between the two sensors in the sensitivity (p=0.08 and 0.06) and specificity (p=0.13) for the diagnosis of resorption in both root areas. The overall accuracy of CCD was higher than PSP sensor; how-ever, the difference was not statistically significance (p>0.05). Conclusion. CCD and PSP sensors chosen for the present study produced similar results in diagnosing simulated external root resorption. |
Databáze: | Directory of Open Access Journals |
Externí odkaz: |