Ex-PRESS implantation for different types of glaucoma
Autor: | Meltem Guzin Altinel, Ayse Yagmur Kanra, Remzi Karadag, Huseyin Bayramlar |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2019 |
Předmět: | |
Zdroj: | International Journal of Ophthalmology, Vol 12, Iss 8, Pp 1290-1297 (2019) |
Druh dokumentu: | article |
ISSN: | 2222-3959 2227-4898 |
DOI: | 10.18240/ijo.2019.08.09 |
Popis: | AIM: To evaluate the clinical results, efficacy and safety of Ex-PRESS P200 glaucoma shunt implant in different types of medically uncontrolled glaucoma. METHODS: The study included 31 eyes of 31 patients that were unresponsive to medical antiglaucomatous therapy in whom Ex-PRESS P200 glaucoma shunt implantation was performed. The demographic characteristics of the patients, type of glaucoma, complete ocular examination results, number of antiglaucomatous drugs before and after surgery, early and late complications of surgery, additional surgical and nonsurgical medical interventions, and success rates were investigated from the patients’ files, retrospectively. RESULTS: The mean postoperative follow-up time was 16.4±7.5mo. The preoperative mean corrected intraocular pressure (IOP) was 28.7±10.3 mm Hg and postoperative mean corrected IOP was 15.3±5.2 mm Hg (P |