Comparison of the Results of Sodium, Potassium, and Chlorine Measured Through Ion Selective Electrode (ISE) Method in Different Devices

Autor: Ali Ünlü, Hüseyin Dost, Saadet Kader, Turan Akdağ
Jazyk: angličtina
Rok vydání: 2022
Předmět:
Zdroj: Genel Tıp Dergisi, Vol 32, Iss 2, Pp 198-202 (2022)
Druh dokumentu: article
ISSN: 2602-3741
DOI: 10.54005/geneltip.1038123
Popis: Objective: The measurements of electrolytes are an indispensable part of the routine tests performed in clinical biochemistry laboratories. In the present study, we aimed to measure serum sodium (Na), potassium (K), and chlorine (Cl) electrolytes via the ion-selective electrode (ISE) method and evaluated in terms of linearity, precision, repeatability, and accuracy levels in the Mindray BS-800M and Beckman Coulter AU5800 devices. Material and methods: The serum samples were obtained from the healthy volunteers admitted to the Karapınar State Hospital for control purposes between January and April 2021. The sample probe was conducted using the same indirect method in the Mindray BS-800M and Beckman Coulter AU5800 devices, which is an individual sample probe for the ISE method. Results: While the comparative measurements were performed on different devices, the correlations between the electrolytes were found as follows: Na [r=0.964, 95% CI (0.52-0.90)], K [r=0.995, 95% CI (0.09-0.12)], and Cl [r=0.972, 95% CI (0.69-1.12)]. Moreover, a correlation was also found between these results. Conclusion: A consistency between the measured Na, K, and Cl electrolytes was determined on the different analyzers. Also, a positive correlation was determined between those electrolytes. We consider that further and comprehensive studies are needed to elucidate these results.
Databáze: Directory of Open Access Journals