Autor: |
Elinor Goshen, Leonid Beilin, Eli Stern, Tal Kenig, Ronen Goldkorn, Simona Ben-Haim |
Jazyk: |
angličtina |
Rok vydání: |
2018 |
Předmět: |
|
Zdroj: |
EJNMMI Physics, Vol 5, Iss 1, Pp 1-12 (2018) |
Druh dokumentu: |
article |
ISSN: |
2197-7364 |
DOI: |
10.1186/s40658-018-0205-z |
Popis: |
Abstract Background The performance of a prototype novel digital single-photon emission computed tomography (SPECT) camera with multiple pixelated CZT detectors and high sensitivity collimators (Digital SPECT; Valiance X12 prototype, Molecular Dynamics) was evaluated in various clinical settings. Images obtained in the prototype system were compared to images from an analog camera fitted with high-resolution collimators. Clinical feasibility, image quality, and diagnostic performance of the prototype were evaluated in 36 SPECT studies in 35 patients including bone (n = 21), brain (n = 5), lung perfusion (n = 3), and parathyroid (n = 3) and one study each of sentinel node and labeled white blood cells. Images were graded on a scale of 1–4 for sharpness, contrast, overall quality, and diagnostic confidence. Results Digital CZT SPECT provided a statistically significant improvement in sharpness and contrast in clinical cases (mean score of 3.79 ± 0.61 vs. 3.26 ± 0.50 and 3.92 ± 0.29 vs. 3.34 ± 0.47 respectively, p |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|
Nepřihlášeným uživatelům se plný text nezobrazuje |
K zobrazení výsledku je třeba se přihlásit.
|