Understanding the Stress Distribution on Anatomic Customized Root-Analog Dental Implant at Bone-Implant Interface for Different Bone Densities

Autor: Pawhat Nimmawitt, Abdul Azeez Abdu Aliyu, Boonrat Lohwongwatana, Sirida Arunjaroensuk, Chedtha Puncreobutr, Nikos Mattheos, Atiphan Pimkhaokham
Jazyk: angličtina
Rok vydání: 2022
Předmět:
Zdroj: Materials, Vol 15, Iss 18, p 6379 (2022)
Druh dokumentu: article
ISSN: 1996-1944
DOI: 10.3390/ma15186379
Popis: The aim of this study is to assess the stress distribution on the bone tissue and bone-implant interface of a customized anatomic root-analog dental implant (RAI) by means of finite element analysis (FEA) for different types of bone density. A mandibular right second premolar was selected from the CBCT database. A DICOM file was converted to an STL file to create a CAD model in FEA software. The bone boundary model was created, while bone density types I–IV were determined. Von Mises stress was measured at bone tissues and bone-implant interfaces. To validate the models, the RAI was 3D printed through a laser powder-bed fusion (L-PBF) approach. The results revealed that all RAI designs could not cause plastic deformation or fracture resulting in lower stress than the ultimate tensile stress of natural bone and implant. Compared to a conventional screw-type implant, RAIs possess a more favorable stress distribution pattern around the bone tissue and the bone-implant interface. The presence of a porous structure was found to reduce the stress at cancellous bone in type IV bone density.
Databáze: Directory of Open Access Journals