Dedicated cardiac rehabilitation wearable sensor and its clinical potential.

Autor: Hooseok Lee, Heewon Chung, Hoon Ko, Changwon Jeong, Se-Eung Noh, Chul Kim, Jinseok Lee
Jazyk: angličtina
Rok vydání: 2017
Předmět:
Zdroj: PLoS ONE, Vol 12, Iss 10, p e0187108 (2017)
Druh dokumentu: article
ISSN: 1932-6203
DOI: 10.1371/journal.pone.0187108
Popis: We describe a wearable sensor developed for cardiac rehabilitation (CR) exercise. To effectively guide CR exercise, the dedicated CR wearable sensor (DCRW) automatically recommends the exercise intensity to the patient by comparing heart rate (HR) measured in real time with a predefined target heart rate zone (THZ) during exercise. The CR exercise includes three periods: pre-exercise, exercise with intensity guidance, and post-exercise. In the pre-exercise period, information such as THZ, exercise type, exercise stage order, and duration of each stage are set up through a smartphone application we developed for iPhones and Android devices. The set-up information is transmitted to the DCRW via Bluetooth communication. In the period of exercise with intensity guidance, the DCRW continuously estimates HR using a reflected pulse signal in the wrist. To achieve accurate HR measurements, we used multichannel photo sensors and increased the chances of acquiring a clean signal. Subsequently, we used singular value decomposition (SVD) for de-noising. For the median and variance of RMSEs in the measured HRs, our proposed method with DCRW provided lower values than those from a single channel-based method and template-based multiple-channel method for the entire exercise stage. In the post-exercise period, the DCRW transmits all the measured HR data to the smartphone application via Bluetooth communication, and the patient can monitor his/her own exercise history.
Databáze: Directory of Open Access Journals