A case of two connected stents deployed during iStent inject W surgery

Autor: Ayaka Shimada, Sho Ichioka, Akiko Ishida, Sachiko Kaidzu, Masaki Tanito
Jazyk: angličtina
Rok vydání: 2023
Předmět:
Zdroj: BMC Ophthalmology, Vol 23, Iss 1, Pp 1-4 (2023)
Druh dokumentu: article
ISSN: 1471-2415
DOI: 10.1186/s12886-023-02951-z
Popis: Abstract Background We report a case with two connected stents ejected simultaneously during an iStent inject W surgery, a modified second-generation iStent Trabecular Micro-Bypass System. Case presentation A 57-year-old woman with primary open-angle glaucoma underwent a combined cataract and iStent inject W surgery in her left eye. After the trabecular meshwork/Schlemm’s canal was pierced by the trocar of injector, the delivery button was pressed a first time, but the stent was not ejected. After the button was pressed a second time, connected two stents were ejected. After removing the dislocated stents from the anterior chamber, two stents were implanted into the desired places using another injector. Except for mild hyphema, no postoperative complication occurred. Stereomicroscopic observation showed that the two stents were connected by a broken trocar shaft. An X-ray showed that the trocar shaft was broken at the part referred to as the “sprayed trocar”. Scanning electron microscopy showed that the surface features of the broken trocar and trocar tip represented tensile failure. Conclusions Although rare, considering that the damage was seen at the structurally weak part (i.e., sprayed trocar), the same phenomenon can happen. For patient safety, surgeons are recommended to inspect the device when the deployment of either the first or second stent is unsuccessful during the iStent inject surgery.
Databáze: Directory of Open Access Journals