HPLC-CUPRAC post-column derivatization method for the determination of antioxidants: a performance comparison between porous silica and core-shell column packing

Autor: Syed A. Haque, Socrates Jose P. Cañete
Jazyk: angličtina
Rok vydání: 2018
Předmět:
Zdroj: Journal of Analytical Science and Technology, Vol 9, Iss 1, Pp 1-9 (2018)
Druh dokumentu: article
ISSN: 2093-3371
DOI: 10.1186/s40543-018-0137-1
Popis: Abstract Background An HPLC method employing a post-column derivatization strategy using the cupric reducing antioxidant capacity reagent (CUPRAC reagent) for the determining antioxidants in plant-based materials leverages the separation capability of regular HPLC approaches while allowing for detection specificity for antioxidants. Methods Three different column types, namely core-shell and porous silica including two chemically different core-shell materials (namely phenyl-hexyl and C18), were evaluated to assess potential improvements that could be attained by changing from a porous silica matrix to a core-shell matrix. Tea extracts were used as sample matrices for the evaluation specifically looking at catechin and epigallocatechin gallate (EGCG). Results Both the C18 and phenyl-hexyl core-shell columns showed better performance compared to the C18 porous silica one in terms of separation, peak shape, and retention time. Among the two core-shell materials, the phenyl-hexyl column showed better resolving power compared to the C18 column. Conclusions The CUPRAC post-column derivatization method can be improved using core-shell columns and suitable for quantifying antioxidants, exemplified by catechin and EGCG, in tea samples.
Databáze: Directory of Open Access Journals
Nepřihlášeným uživatelům se plný text nezobrazuje