Radiological Shielding Design for the Neutron High-Resolution Backscattering Spectrometer EMU at the OPAL Reactor

Autor: Ersez Tunay, Esposto Fernando, Souza Nicolas R. de
Jazyk: angličtina
Rok vydání: 2017
Předmět:
Zdroj: EPJ Web of Conferences, Vol 153, p 05002 (2017)
Druh dokumentu: article
ISSN: 2100-014X
DOI: 10.1051/epjconf/201715305002
Popis: The shielding for the neutron high-resolution backscattering spectrometer (EMU) located at the OPAL reactor (ANSTO) was designed using the Monte Carlo code MCNP 5-1.60. The proposed shielding design has produced compact shielding assemblies, such as the neutron pre-monochromator bunker with sliding cylindrical block shields to accommodate a range of neutron take-off angles, and in the experimental area - shielding of neutron focusing guides, choppers, flight tube, backscattering monochromator, and additional shielding elements inside the Scattering Tank. These shielding assemblies meet safety and engineering requirements and cost constraints. The neutron dose rates around the EMU instrument were reduced to < 0.5 µSv/h and the gamma dose rates to a safe working level of ≤ 3 µSv/h.
Databáze: Directory of Open Access Journals