Finite Element Analysis of Bone Stress around Micro-Implants of Different Diameters and Lengths with Application of a Single or Composite Torque Force.

Autor: Ying-juan Lu, Shao-hai Chang, Jian-tao Ye, Yu-shan Ye, Yan-song Yu
Jazyk: angličtina
Rok vydání: 2015
Předmět:
Zdroj: PLoS ONE, Vol 10, Iss 12, p e0144744 (2015)
Druh dokumentu: article
ISSN: 1932-6203
DOI: 10.1371/journal.pone.0144744
Popis: BackgroundStress on the bone surrounding dental micro-implants affects implant success.PurposeTo compare the stress on the bone surrounding a micro-implant after application of a single force (SF) of 200 g or a composite force (CF) of 200 g and 6 N.mm torque.Materials and methodsFinite element models were developed for micro-implant diameters of 1.2, 1.6, and 2.0 mm, and lengths of 6, 8, 10, and 12 mm and either a SF or CF was applied. The maximum equivalent stress (Max EQS) of the bone surrounding the micro-implant was determined, and the relationships among type of force, diameter, and length were evaluated.ResultsThe Max EQS of the CF exceeded that of the SF (P< 0.05). The effect of force on stress was related to implant diameter, but not to implant length. The larger CF led to greater instability of the micro-implant and the effect was most pronounced at an implant diameter of 1.2 mm. The use of implant diameters of 1.6 mm and 2.0 mm produced no significant difference in implant stability when either a CF or SF was applied.ConclusionWhen considering the use of an implant to perform three-dimensional control on the teeth, the implant diameter chosen should be > 1.2 mm.